Instrument features
  • Spectral response ....................... 8 - 14 μm
  • Optical .......................................... D.S : 50:1 ( 50.8 mm at 2540 mm )
  • Temperature range ...................... From -50 to +1850 C
  • Accuracy .................................... From -50 to +20 C : ±3 C
  •                                                            From +20 to +500 C : ±1% ±1 C

                                                               From +500 to +1000 C : ±1.5%

                                                               From +1000 to +1850 C : ±2%
  • Infrared repeatability ................... From -50 to +20 C : ±1.5 C
  •                                                              From +20 to +1000 C : ±0.5% or ±0.5 C

                                                                   From +1000 to +1850 C : ±1%
  • Display resolution ....................... 0.1 C
  • Response time ............................. 150 ms
  • Emissivity .................................... Adjustable from 0.10 to 1.0 (pre-set at 0.95)
  • Over range indication ................ Display indication : « ---- »
  • Dual laser sighting ...................... Wavelength : from 630 nm to 670 nm Output < 1mW, Class 2 (II)
  • Positive or negative
  • temperature indication ................ Automatic ( no indication for a positive temperature ) (-) sign for a negative temperature
  • Display ......................................... 3 lines, 4 digits with back lighted display LCD
  • Auto-extinction ............................ Automatic after 7 seconds of inactivity
  • High/low alarm ............................. Flashing signal on display and beep signal with adjustable thresholds
  • Power supply ............................... Alkaline 9V battery
  • Autonomy .................................... 95 h (inactive laser and backlight)
  •                                                              15 h (active laser and backlight)
  • Use temperature .......................... From 0 to +10 C for a short period
  •                                                                From +11 to +50 C for a long period
  • Storage temperature ................... From -10 C to +60 C
  • Relative humidity ........................ From 10 to 90%RH in operating mode and lower than 80%RH in storage
  • Dimensions .................................. 200 x 140 x 50 mm
  • Weight .......................................... 320 g (included battery)
  • Memory ........................................ 100 temperature values
  • Accuracy for an ambient temperature from 23 to 25 C (with a relative humidity lower than 80% RH)

Thermocouple K probe features
  • Temperature range ...................... From -40 to 400 C
  • Display range ............................... From -50 to +1370 C
  • Resolution ................................... 0.1 C
  • Accuracy ...................................... ±1.5% of reading ±3 C
  • Cable length ................................ 1 m